Air quality data is used for compliance, for research and increasingly for public reporting. In all three cases the value of a measurement depends on the method behind it. Two instruments can report the same number for the same air and be worth very different amounts, because one is a reference method and the other is an indicative sensor with an unknown correction.
Reference methods and indicative methods
A reference method is defined by a standard — for particulate matter, gravimetric sampling onto a filter; for gases, a specified analytical technique such as chemiluminescence for NOx, UV fluorescence for SO₂, or non-dispersive infrared for CO. Reference methods are slow, expensive and legally defensible.
Indicative methods are faster and cheaper: optical particle counters, electrochemical gas sensors, metal-oxide sensors. They are useful for screening, for spatial density and for trend monitoring, but they require co-location calibration against a reference method before their numbers can be compared with a limit value.
The practical question is not "which is better" but "what does my data need to support". A compliance report needs a reference or formally validated equivalent method. A survey of fifteen sites needs indicative instruments, co-located against a reference, with the correction documented.
Particulates: the size fraction is the measurement
PM10, PM2.5 and PM1 are not the same pollutant at different resolutions. They are defined size fractions with different sources, different health effects and different limit values. An optical sensor reports a mass concentration derived from scattering, calibrated against a specific aerosol; its response to humidity, to smoke, and to coarse dust differs from a gravimetric reference.
- Dry the sample or correct for humidity; hygroscopic growth inflates optical readings dramatically above about 60% relative humidity.
- Check the calibration aerosol. A sensor calibrated on Arizona test dust will not report diesel soot correctly.
- Co-locate for at least a week at each distinct season or source regime before using indicative data quantitatively.
- Report the method alongside the number, always.
Gases: cross-interference is the hidden error
Electrochemical sensors are affordable and sensitive, and they cross-react. An NO₂ sensor may respond to ozone; an H₂S sensor may respond to SO₂; a CO sensor may respond to hydrogen. For regulatory work, the analyser should use a technique with defined selectivity: chemiluminescence for NOx, UV fluorescence for SO2, NDIR or gas filter correlation for CO, UV photometry for ozone.
Calibration gases must be traceable and appropriate to the range: a cylinder at 80% of the analyser's full scale gives a meaningful span check, while a cylinder at 5% does not. Zero air matters as much as span gas.
Siting decides more than the instrument
A monitor in the wake of a building, next to an air-conditioning exhaust, or under a tree measures its own micro-environment. Reference siting guidance sets minimum distances from walls, roads and obstructions, and a defined inlet height — typically 1.5 to 4 m above ground for ambient monitoring, with the inlet clear of nearby surfaces.
For indoor monitoring, the question is different: where do people breathe? A sensor on a shelf near the ceiling reports a different number from one at desk height, and the difference is not noise.
Calibration, maintenance and data capture
- Zero and span checks on a defined schedule, with the results recorded.
- Filter and inlet maintenance, because a loaded filter changes the sample flow and the response time.
- Flow verification: many gas analysers assume a stable sample flow, and a pump that is fading biases every reading.
- Data capture targets: a compliance dataset usually requires 75-90% valid data; plan maintenance so it does not eat into that.
- Metadata: instrument serial number, firmware, calibration date, siting coordinates and inlet height travel with every dataset.
Choosing an instrument for a laboratory
Laboratories monitoring workplace air, fume hood performance, or emissions from a process need a different specification from an ambient network. State the parameter, the range, the limit value, the required method, the data capture requirement and the reporting format. Then choose an instrument that meets the method, not one that meets the price.