An autoclave is often bought by volume โ 18 litres, 23 litres, 50 litres โ when the decision that actually determines whether a load comes out sterile is the sterilizer class. Class B, Class N and Class S machines look similar on the outside. Inside, they differ in whether air can be removed from the load before steam is admitted, and that single difference decides which loads they can process safely.
Why air removal is the whole story
Steam sterilization works by condensing saturated steam on the load and releasing latent heat. If air remains inside a wrapped pack, a hollow instrument or a porous load, the steam cannot reach the surface, and the temperature that the chamber sensor reports is not the temperature the microorganisms experience. Air is also denser than steam and collects in the coldest part of the chamber, which is usually the drain.
Every autoclave class is really an answer to the question: how is air removed before sterilization starts?
Class N: no vacuum, unwrapped solid items only
Class N sterilizers remove air by displacement โ steam enters at the top and pushes air out through a valve. There is no vacuum pump. This works acceptably for unwrapped solid instruments, glassware, and liquids in vented containers, provided the load is not tightly packed.
It does not work for wrapped instruments, hollow or lumen devices, porous loads, or textiles, because steam cannot displace air trapped inside those geometries. A Class N machine sold as a general-purpose laboratory autoclave will eventually be used for a wrapped load, and that is where sterilization failures come from.
Class B: fractionated vacuum, the full envelope
Class B sterilizers use a vacuum pump to evacuate the chamber, admit steam, and repeat โ the fractionated pre-vacuum cycle. Removing air by vacuum rather than displacement reaches into wrapped packs, hollow instruments, porous loads, textiles and implant trays. Class B is the class required by EN 13060 for the widest range of loads, and it is the class that matches most hospital, dental and laboratory expectations.
The vacuum stage also enables the leak test that verifies chamber integrity, and post-cycle vacuum drying, which matters when instruments must be stored sterile rather than used immediately.
Class S: a defined subset, and why the definition matters
Class S sits between the two. The manufacturer specifies exactly which loads the sterilizer is validated for โ typically unwrapped solid items, and perhaps single-wrapped instruments or specific porous loads. The machine may use a partial vacuum, a supervised air removal stage, or a particular cycle design.
The practical consequence: a Class S autoclave is not a cheaper Class B. It is a machine with a declared, narrower load list. If your load list changes later โ you start wrapping instruments, or add lumened devices โ a Class S machine may no longer be suitable even though it is working exactly as designed.
Matching the class to the loads you actually process
- Unwrapped solid instruments, glassware, media preparation: Class N is adequate, and is common in small laboratories and teaching facilities.
- Wrapped instruments, hollow or lumened devices, porous loads, textiles: Class B. This is the default for hospitals, dental clinics, veterinary practices and busy laboratories.
- A declared, stable load list that matches a manufacturer specification: Class S may be acceptable, but record the load list and revisit it whenever practice changes.
What else the class does not tell you
Two machines of the same class can differ substantially in chamber volume, cycle time, water consumption, door design, documentation and serviceability. A Class B autoclave with a 3.5 litre clean and waste water tank is a different proposition from one that must be plumbed in. Look at:
- Cycle records: USB export, printer, or network logging, and whether records are retained.
- Programs: solid, packaged, fabric, liquids, B&D test, helix test, vacuum leak test.
- Water: tank volume and whether clean and waste water are separate.
- Chamber material and finish: S30408 stainless steel is typical; welds and finish affect cleanability.
- Consumables and service: door gaskets, filters, and who performs the annual validation.
A note on validation, whatever the class
Class describes the air removal method, not the quality of the installation. A Class B machine with a saturated filter, a worn door gasket or an uncalibrated sensor will fail loads. Daily Bowie-Dick or helix tests, a weekly vacuum leak test where the machine supports it, and an annual calibration and validation visit are what keep a sterilizer trustworthy. If a supplier cannot explain how the machine will be validated and by whom, the class label is not worth much.
Krivalis supplies Class B autoclaves in 18 L and 23 L chamber volumes with programmable vacuum and sterilization cycles, USB data export, 0.22 ยตm air filtration and a separate clean and waste water tank. If you are unsure which class your load list requires, send the list โ wrapped, unwrapped, hollow, porous, liquids โ and we will confirm the configuration. The autoclave buying guide covers sizing and validation in more detail.